首页    期刊浏览 2024年11月30日 星期六
登录注册

文章基本信息

  • 标题:Micromorphometric change of TiO2 b last i mplant surface conditioned with tetracycline-HCl
  • 本地全文:下载
  • 作者:Lee, Jung Min ; Herr, Yeek ; Kwon, Young Hyuk
  • 期刊名称:The Journal of the Korean Academy of Periodontology
  • 印刷版ISSN:0250-3352
  • 出版年度:2007
  • 卷号:37
  • 期号:3
  • 页码:599-611
  • DOI:10.5051/jkape.2007.37.3.599
  • 语种:Korean
  • 出版社:Korean Academy of Periodontology
  • 摘要:

    The present study was performed to evaluate the surface roughness and effect of Tetracycline-HCl on the change of implant surface microstructure according to application time. TiO2 surface Implant was utilized. Implant surface was rubbed with 50mg/ml Tetracycline-HCl solution for 0.5min, 1min, 1.5min, 2min, 2.5min and 3min respectively in the Tetracycline-HCl group. Then, specimens were measured surface roughness and processed for scanning electron microscopic observation.

    The results of this study were as follows.

    TiO2 blast implant surface showed increased surface roughness 1.5 minute after treatment with Tetracycline-HCl. But, there were not significant differences in saline group after treatment.

    Tetracycline-HCl group showed changed surface micro-morphology in SEM after 1.5 minute. There were not significant differences in saline group after treatment.

    Between Tetracycline-HCl group and saline group, there were difference in surface roughness change and SEM micro-morphology.

    Tetracycline-HCl have influence on TiO2 blast implant surface.

    In conclusion, the detoxification with 50mg/ml Tetracycline-HCl must be applied respectively with different time according to various implant surfaces.

  • 关键词:TiO2 blast implant surface; tetracycline-HCl; Detoxification
国家哲学社会科学文献中心版权所有