  • 标题:Corneal Thickness Measured by Dual Scheimpflug, Anterior Segment Optical Coherence Tomography, and Ultrasound Pachymetry
  • 本地全文:下载
  • 作者:Kim, Dong Wook ; Yi, Ka Young ; Choi, Dong Gyu
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2012
  • 卷号:53
  • 期号:10
  • 页码:1412-1418
  • DOI:10.3341/jkos.2012.53.10.1412
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare central corneal thickness (CCT) as measured by dual rotating Scheimpflug camera (Galilei), anterior segment optical coherence tomography (AS-OCT), and ultrasound pachymetry (USP).

    Methods

    The measurements of CCT using a dual rotating Scheimpflug camera, AS-OCT, and USP in 40 eyes of 20 healthy subjects were compared.

    Results

    The average measurements of CCT by dual rotating Scheimpflug camera, AS-OCT, and USP were 538.10 ± 31.36 µm, 536.20 ± 31.21 µm, and 541.93 ± 34.93 µm, respectively. The CCT measurement by USP was statistically significantly thicker than by the dual rotating Scheimpflug camera and AS-OCT ( p = 0.017, p = 0.001, respectively). There was no significant difference between the dual rotating Scheimpflug camera and AS-OCT ( p = 0.054). A significant linear correlation was observed between the dual rotating Scheimpflug camera, the AS-OCT, and the USP (r > 0.900, p < 0.001).

    Conclusions

    The results of the 3 methods have significant correlation with each other, but the measurement by USP was significantly thicker than the dual rotating Scheimpflug camera and AS-OCT. Therefore, CCT should be interpreted in the context of the instrument used.

  • 关键词:Anterior segment optical coherence tomography; Central corneal thickness; Dual rotating scheimpflug camera; GALILEI; Ultrasound pachymetry