首页    期刊浏览 2024年11月30日 星期六
登录注册

文章基本信息

  • 标题:Comparison of Electroretinogram Waveforms Acquired Using Monopolar ERG-Jet Electrode and Bipolar Burian-Allen Electrode
  • 本地全文:下载
  • 作者:Park, Jihyun ; Lee, Hokyung ; Park, Sung Joon
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2010
  • 卷号:51
  • 期号:9
  • 页码:1217-1223
  • DOI:10.3341/jkos.2010.51.9.1217
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare electroretinogram (ERG) waveforms acquired using an ERG-jet electrode and a Burian-Allen electrode.

    Methods

    ERGs were recorded with an ERG-jet electrode and a Burian-Allen electrode from both eyes of 29 volunteers. Three consecutive recordings were obtained with both electrodes from the left eye of another 6 volunteers. Peak-to-trough amplitudes and peak implicit times were compared between both eyes, and between the 2 types of electrodes. Interpersonal and intrapersonal variation were also compared.

    Results

    The differences in amplitude and implicit time between the right and left eyes were not influenced by the type of electrode. The amplitude of the ERG acquired using the Burian-Allen electrode, however, showed significant differences between both eyes, contrary to the amplitude acquired using the ERG-jet electrode. ERG recordings obtained using the ERG-jet electrode generally showed higher amplitude, shorter implicit time, and less interpersonal and intrapersonal variation than ERG recordings obtained using the Burian-Allen electrode.

    Conclusions

    In this application, the ERG-jet ERG electrode appeared to be superior to the Burian-Allen electrode in terms of both usefulness and consistency.

  • 关键词:Burian-Allen electrode; Electroretinogram; ERG-Jet electrode
国家哲学社会科学文献中心版权所有