首页    期刊浏览 2024年12月12日 星期四
登录注册

文章基本信息

  • 标题:Comparison of Time Domain OCT and Spectrum Domain OCT for Retinal Nerve Fiber Layer Assessment
  • 本地全文:下载
  • 作者:Kim, Bu Ki ; Lee, Dong Wook ; Ahn, Min
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2009
  • 卷号:50
  • 期号:10
  • 页码:1539-1547
  • DOI:10.3341/jkos.2009.50.10.1539
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare retinal nerve fiber layer (RNFL) thickness obtained with Stratus optical coherence tomography (OCT) and Cirrus OCT.

    Methods

    Sixty-one normal eyes were evaluated with Stratus and Cirrus OCT on the same day, and the RNFL thicknesses measured by the two OCT machines were compared. The correlation between the two data sets was obtained using Pearson's correlation coefficient. The correlation between RNFL thickness and the difference in data measured by the two OCT machines was then assessed.

    Results

    The average RNFL thickness was significantly higher with Stratus OCT by 6.54±4.48 µm (p=0.0008). A strong correlation was present between the two RNFL thickness data sets (r=0.883), and the difference between Stratus and Cirrus values tended to increase as RNFL thickness increased.

    Conclusions

    RNFL thickness measurements in normal eyes scanned with Cirrus OCT correlate well with Stratus OCT measurements. Average RNFL thickness was significantly higher with Stratus OCT, and as the RNFL thickness increased, the difference between Stratus and Cirrus values increased.

  • 关键词:Cirrus; Optical coherence tomography; Retinal nerve fiber layer thickness; Stratus