  • 标题:Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer -A case report-
  • 本地全文:下载
  • 作者:Kim, Hyo Jin ; Kim, Mi Woon
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2010
  • 卷号:59
  • 期号:4
  • 页码:270-274
  • DOI:10.4097/kjae.2010.59.4.270
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000® vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we couldn't check if he had experienced awareness with recall. We emphasize the importance of checking the anesthetic vaporizer after mounting it on the back bar of the anesthesia machine.

  • 关键词:Gas leak; Selectatec; Vaporizer