首页    期刊浏览 2024年12月04日 星期三
登录注册

文章基本信息

  • 标题:The effect of ultrasonic cleaning on the elimination of postoperative residual protein from the proseal™ laryngeal mask airway
  • 本地全文:下载
  • 作者:Choi, Won-Jun ; Kim, Yun Hong ; Kim, Hyun Soo
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2008
  • 卷号:55
  • 期号:4
  • 页码:424-428
  • DOI:10.4097/kjae.2008.55.4.424
  • 语种:Korean
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:Background

    Reusable Proseal™ laryngeal mask airways (PLMAs) can act as a vector for the transmission of prion diseases such as variant Creutzfeldt-Jacob disease. This study tested the hypothesis that supplementary ultrasonic cleaning facilitates the removal of protein deposits on PLMAs after anesthesia.

    Methods

    After clinical use, 40 PLMAs were randomly allocated into two groups. In the first group, the PLMAs were washed by hand and were then subsequently placed in an autoclave at 134℃ for 40 min (Group 1, n = 20). In the second group, the PLMAs were washed by hand and ultrasonic cleaning using an enzymatic solution for 5 min, and were then subsequently placed in an autoclave (Group 2, n = 20). In both groups, protein deposits were detected on PLMAs by erythrosin staining. A staining score designated as none (0%), mild (0-20%), moderate (20-80%) and severe (80-100%), was assigned to each site (outer surface, inner surface and edges of the cuff, airway and drain tube, finger strap) according to the percentage of the stained surface area.

    Results

    Despite the cleaning of the masks, residual protein was found on the outer surface, inner surface and edge of the cuff, airway and drain tube, and finger strap of the PLMAs in both groups. Similar scores were observed for each part of the cleaned PLMAs in both groups, except for the outer surface of the PLMAs in Group 2 (P < 0.05).

    Conclusions

    We conclude that the use of an ultrasonic cleaner with an enzymatic solution may be effective to cleanse the outer surface of the PLMAs, but there were no differences in the total scores for both groups.

  • 关键词:erythrosin; laryngeal mask; proteins; ultrasonic
国家哲学社会科学文献中心版权所有