  • 标题:Comparative evaluation of the accuracy of linear measurements betweencone beam computed tomographyand 3D microtomography
  • 本地全文:下载
  • 作者:Francesca Mangione ; Deborah Meleo ; Marco Talocco
  • 期刊名称:Annali dell'Istituto Superiore di Sanita
  • 印刷版ISSN:0021-2571
  • 出版年度:2013
  • 卷号:49
  • 期号:3
  • 页码:261-265
  • DOI:10.4415/ANN_13_03_05
  • 出版社:Editrice Kurtis Srl
  • 摘要:Objective. The aim of this study was to evaluate the influence of artifacts on the accuracy of linear measurements estimated with a common cone beam computed tomography (CBCT) system used in dental clinical practice, by comparing it with microCT system as standard reference. Materials and Methods. Ten bovine bone cylindrical samples containing one implant each, able to provide both points of reference and image quality degradation, have been scanned by CBCT and microCT systems. Thanks to the software of the two systems, for each cylindrical sample, two diameters taken at different levels, by using implants different points as references, have been measured. Results have been analyzed by ANOVA and a significant statistically difference has been found. Results and Discussion. Due to the obtained results, in this work it is possible to say that the measurements made with the two different instruments are still not statistically comparable, although in some samples were obtained similar performances and therefore not statistically significant. Conclusion. With the improvement of the hardware and software of CBCT systems, in the near future the two instruments will be able to provide similar performances
  • 关键词:cone-beam CT;. microCT;. linear measurements;. artifacts