首页    期刊浏览 2024年12月12日 星期四
登录注册

文章基本信息

  • 标题:Successive gamma-ray irradiation and corresponding post-irradiation annealing of pMOS dosimeters
  • 本地全文:下载
  • 作者:Pejović Milić M. ; Pejović Momčilo M. ; Jakšić Aleksandar B.
  • 期刊名称:Nuclear Technology and Radiation Protection
  • 印刷版ISSN:1451-3994
  • 出版年度:2012
  • 卷号:27
  • 期号:4
  • 页码:341-345
  • DOI:10.2298/NTRP1204341P
  • 出版社:VINČA Institute of Nuclear Sciences
  • 摘要:

    The paper investigates a possibility of pMOS dosimeter re-use for the measurement of gamma-ray irradiation. The dosimeters were irradiated to the dose of 35 Gy, annealed at room and elevated temperatures, after which they were irradiated again to the same dose value. Changes in the threshold voltage shift during those processes were followed, and it was shown that their re-use depends on a gate polarization during irradiation. For the gate polarization of 5 V during irradiation the pMOS dosimeters can be re-used for measurements of the irradiation dose after annealing without prior calibration. The pMOS dosimeters with the gate polarization during irradiation of 2.5 V can also be re-used for irradiation dose measurements but they require calibration. It is shown that for their re-use it is necessary to anneal the pMOS dosimeter so that the fading is higher than 50%. [Projekat Ministarstva nauke Republike Srbije, br. 171007]

  • 关键词:pMOS dosimeter; gamma-ray irradiation; threshold voltage shift; absorbed dose
国家哲学社会科学文献中心版权所有