  • 标题:A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring: Pilot Test in a Pig Model
  • 本地全文:下载
  • 作者:Joost Wauters ; Liesbeth Spincemaille ; Anne-Sophie Dieudonne
  • 期刊名称:Critical Care Research and Practice
  • 印刷版ISSN:2090-1305
  • 电子版ISSN:2090-1313
  • 出版年度:2012
  • 卷号:2012
  • DOI:10.1155/2012/181563
  • 出版社:Hindawi Publishing Corporation
  • 摘要:Background. Intravesical pressure (IAPivp) measurement is considered to be the gold standard for assessment of intra-abdominal pressure (IAP). This study evaluated a new minimally invasive IAP monitoring device (CiMON) against three other devices in a wide range of clinically relevant IAP and in different body positions in healthy pigs. Methods. The CiMON catheter (IAPCiM) and another balloon-tipped catheter (IAPspie) were positioned into the stomach. Fluid-filled catheters were used for direct intraperitoneal (IAPdir) and IAPivp measurement. Both in supine and 25∘ head-of-bed positions, IAP was increased from baseline to 30 mmHg. At every IAP level, 4 IAP measurements were recorded simultaneously. Mean differences and the limits of agreement were calculated. Results. Bias between IAPCiM and IAPspie was nearly zero with very good agreement, both in supine and 25∘ position. In supine position, IAPCiM slightly overestimated IAPivp and IAPdir by 1.5 and 2.1 mmHg with reasonable agreement. In 25∘ position, IAPCiM underestimated IAPivp and IAPdir by 1.0 and 0.5 mmHg, again with reasonable agreement. Conclusions. Agreement between IAPCiM and IAPspie was very good, while good-to-moderate agreement exists between IAPCiM and IAPdir or IAPivp. Simplicity, continuous monitoring, and the combination with a feeding tube should lead to further clinical studies, evaluating this new CiMON device.
国家哲学社会科学文献中心版权所有