摘要:Abstract Radiolucent carbon-fiber-reinforced (CFR) polyethyl-ether-ether-ketone (PEEK) has been established in spinal instrumentation for oncological reasons. Laboratory data reported comparable bacterial adhesion as titanium. Thus, using of CFR-PEEK spinal instrumentation for spondylodiscitis bases on artifact-free imaging to evaluate therapeutic success. Studies comparing the rate of pedicle screw loosening and relapse of spondylodiscitis following titanium versus CFR-PEEK instrumentation do not exist so far. This study evaluates the rate of pedicle screw loosening and recurrence of spondylodiscitis after CFR-PEEK instrumentation for spondylodiscitis compared to titanium. We conducted a prospective single center study between June 2018 and March 2019 on consecutive 23 patients with thoracolumbar spondylodiscitis. Imaging data was evaluated for screw loosening at a minimum of three months after surgery. A matched-pair analysis was performed using spondylodiscitis cases between 2014 and 2016 using titanium instrumentation for equal localization, surgery, and microorganism class. Among 17 cases with follow-up imaging, six cases (35%) showed screw loosening while only 14% (two patients) with titanium instrumentation were loosened ( p = 0.004). In both groups the most frequent bacterium was Staphylococcus aureus , followed by Staphylococcus epidermidis . From the S. aureus cases, one infection in both groups was caused by methicillin resistant species (MRSA). No difference was found in the rate of 360° fusion in either group due to matching criteria. As opposed to other indications CFR-PEEK screws show more loosening than titanium in this series with two potentially underlying reasons: a probably stronger bacterial adhesion on CFR-PEEK in vivo as shown by a statistical trend in vitro and instrumentation of spondylytic vertebrae. Until these factors are validated, we advise caution when implanting CFR-PEEK screws in infectious cases.
其他摘要:Abstract Radiolucent carbon-fiber-reinforced (CFR) polyethyl-ether-ether-ketone (PEEK) has been established in spinal instrumentation for oncological reasons. Laboratory data reported comparable bacterial adhesion as titanium. Thus, using of CFR-PEEK spinal instrumentation for spondylodiscitis bases on artifact-free imaging to evaluate therapeutic success. Studies comparing the rate of pedicle screw loosening and relapse of spondylodiscitis following titanium versus CFR-PEEK instrumentation do not exist so far. This study evaluates the rate of pedicle screw loosening and recurrence of spondylodiscitis after CFR-PEEK instrumentation for spondylodiscitis compared to titanium. We conducted a prospective single center study between June 2018 and March 2019 on consecutive 23 patients with thoracolumbar spondylodiscitis. Imaging data was evaluated for screw loosening at a minimum of three months after surgery. A matched-pair analysis was performed using spondylodiscitis cases between 2014 and 2016 using titanium instrumentation for equal localization, surgery, and microorganism class. Among 17 cases with follow-up imaging, six cases (35%) showed screw loosening while only 14% (two patients) with titanium instrumentation were loosened ( p = 0.004). In both groups the most frequent bacterium was Staphylococcus aureus , followed by Staphylococcus epidermidis . From the S. aureus cases, one infection in both groups was caused by methicillin resistant species (MRSA). No difference was found in the rate of 360° fusion in either group due to matching criteria. As opposed to other indications CFR-PEEK screws show more loosening than titanium in this series with two potentially underlying reasons: a probably stronger bacterial adhesion on CFR-PEEK in vivo as shown by a statistical trend in vitro and instrumentation of spondylytic vertebrae. Until these factors are validated, we advise caution when implanting CFR-PEEK screws in infectious cases.