  • 标题:Accuracy of zero-heat-flux thermometry and bladder temperature measurement in critically ill patients
  • 本地全文:下载
  • 作者:Anselm Bräuer ; Albulena Fazliu ; Thorsten Perl
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2020
  • 卷号:10
  • 期号:1
  • 页码:1-7
  • DOI:10.1038/s41598-020-78753-w
  • 出版社:Springer Nature
  • 摘要:Core temperature (TCore) monitoring is essential in intensive care medicine. Bladder temperature is the standard of care in many institutions, but not possible in all patients. We therefore compared core temperature measured with a zero-heat flux thermometer (TZHF) and with a bladder catheter (TBladder) against blood temperature (TBlood) as a gold standard in 50 critically ill patients in a prospective, observational study. Every 30 min TBlood, TBladder and TZHF were documented simultaneously. Bland–Altman statistics were used for interpretation. 7018 pairs of measurements for the comparison of TBlood with TZHF and 7265 pairs of measurements for the comparison of TBlood with TBladder could be used. TBladder represented TBlood more accurate than TZHF. In the Bland Altman analyses the bias was smaller (0.05 °C vs. − 0.12 °C) and limits of agreement were narrower (0.64 °C to − 0.54 °C vs. 0.51 °C to – 0.76 °C), but not in clinically meaningful amounts. In conclusion the results for zero-heat-flux and bladder temperatures were virtually identical within about a tenth of a degree, although TZHF tended to underestimate TBlood. Therefore, either is suitable for clinical use. German Clinical Trials Register, DRKS00015482, Registered on 20th September 2018, http://apps.who.int/trialsearch/Trial2.aspx?TrialID=DRKS00015482 .
  • 其他摘要:Abstract Core temperature (T Core ) monitoring is essential in intensive care medicine. Bladder temperature is the standard of care in many institutions, but not possible in all patients. We therefore compared core temperature measured with a zero-heat flux thermometer (T ZHF ) and with a bladder catheter (T Bladder ) against blood temperature (T Blood ) as a gold standard in 50 critically ill patients in a prospective, observational study. Every 30 min T Blood , T Bladder and T ZHF were documented simultaneously. Bland–Altman statistics were used for interpretation. 7018 pairs of measurements for the comparison of T Blood with T ZHF and 7265 pairs of measurements for the comparison of T Blood with T Bladder could be used. T Bladder represented T Blood more accurate than T ZHF . In the Bland Altman analyses the bias was smaller (0.05 °C vs. − 0.12 °C) and limits of agreement were narrower (0.64 °C to − 0.54 °C vs. 0.51 °C to – 0.76 °C), but not in clinically meaningful amounts. In conclusion the results for zero-heat-flux and bladder temperatures were virtually identical within about a tenth of a degree, although T ZHF tended to underestimate T Blood . Therefore, either is suitable for clinical use. German Clinical Trials Register, DRKS00015482, Registered on 20th September 2018, http://apps.who.int/trialsearch/Trial2.aspx?TrialID=DRKS00015482 .
国家哲学社会科学文献中心版权所有