摘要:The aim of this study was to explore the reliability and precision of body compartment measures, in particular visceral adipose tissue, in weight stable adults over a range of BMIs using GE-Lunar iDXA. Weight-stable participants aged 18–65 years had a total body composition scan on GE-Lunar iDXA either on three separate occasions over a three month period (n = 51), or on a single occasion for duplicate scans with repositioning (n = 30). The coefficient of variation (CV%) and least significant change (LSC) of body compartments were calculated. The CV was higher for all measures over three months (range 0.8–5.9%) compared with same-day precision-scans (all 500 g. Changes observed in VAT mass levels below 500 g should be interpreted with caution due to lack of precision and reliability. All other compartmental measures demonstrated good reliability, with less than 6% variation over three months.
关键词:adiposity; body composition; dual X-ray absorptiometry; visceral adipose tissue