  • 标题:Toward robust deconvolution of pass-through paleomagnetic measurements: new tool to estimate magnetometer sensor response and laser interferometry of sample positioning accuracy
  • 本地全文:下载
  • 作者:Hirokuni Oda ; Chuang Xuan ; Yuhji Yamamoto
  • 期刊名称:Earth, Planets and Space
  • 电子版ISSN:1880-5981
  • 出版年度:2016
  • 卷号:68
  • 期号:1
  • 页码:1-13
  • DOI:10.1186/s40623-016-0493-2
  • 出版社:Springer Verlag
  • 摘要:Pass-through superconducting rock magnetometers (SRM) offer rapid and high-precision remanence measurements for continuous samples that are essential for modern paleomagnetism studies.However, continuous SRM measurements are inevitably smoothed and distorted due to the convolution effect of SRM sensor response.Deconvolution is necessary to restore accurate magnetization from pass-through SRM data, and robust deconvolution requires reliable estimate of SRM sensor response as well as understanding of uncertainties associated with the SRM measurement system.In this paper, we use the SRM at Kochi Core Center (KCC), Japan, as an example to introduce new tool and procedure for accurate and efficient estimate of SRM sensor response.To quantify uncertainties associated with the SRM measurement due to track positioning errors and test their effects on deconvolution, we employed laser interferometry for precise monitoring of track positions both with and without placing a u-channel sample on the SRM tray.The acquired KCC SRM sensor response shows significant cross-term of Z-axis magnetization on the X-axis pick-up coil and full widths of ~46–54 mm at half-maximum response for the three pick-up coils, which are significantly narrower than those (~73–80 mm) for the liquid He-free SRM at Oregon State University.Laser interferometry measurements on the KCC SRM tracking system indicate positioning uncertainties of ~0.1–0.2 and ~0.5 mm for tracking with and without u-channel sample on the tray, respectively.Positioning errors appear to have reproducible components of up to ~0.5 mm possibly due to patterns or damages on tray surface or rope used for the tracking system.Deconvolution of 50,000 simulated measurement data with realistic error introduced based on the position uncertainties indicates that although the SRM tracking system has recognizable positioning uncertainties, they do not significantly debilitate the use of deconvolution to accurately restore high-resolution signal.The simulated “excursion” event associated with a significant magnetization intensity drop was clearly recovered in the deconvolved measurements with a maximum error of ~3° in inclination.
  • 关键词:Superconducting rock magnetometer; Pass-through measurement; Sensor response; Deconvolution; Laser interferometry; Geomagnetic excursion; U-channel; Sediment long-core sample
国家哲学社会科学文献中心版权所有