  • 摘要:Scleral indentation is widely used to examine the peripheral fundus, however it can increase the intraocular pressure (IOP) to high levels which can then affect retinal function. We evaluated the effects of scleral indentation on the macular function electrophysiologically. Intraoperative focal macular electroretinograms (iFMERGs) were recorded with and without controlling the IOP in 7 eyes. Without IOP control, the IOP increased from 21.7 ± 4.9 to 92.7 ± 20.2 mmHg significantly (P = 0.020) and the amplitudes of the b-wave (from 6.29 ± 1.160 to 3.71 ± 1.98 uV, P = 0.007), on-photopic negative response (from 2.29 ± 0.99 to 0.72 ± 0.47 uV, on-PhNR, P = 0.005), and d-wave (from 2.57 ± 0.41 to 1.64 ± 0.69 uV, P = 0.007) decreased significantly soon after beginning the indentation. All values returned to the baseline levels after releasing the indentation. In the eyes with IOP controlled, the IOP and the amplitude of all components did not change significantly during and after the indentation except the on-PhNR amplitude which was significantly reduced during the indentation. The changes in the iFMERGs and macular function caused by scleral indentation were transient and reversible. The changes can be minimized by controlling the IOP.