  • 标题:Comparison of Tunable Diode Laser Absorption Spectroscopy and Isothermal Micro-calorimetry for Non-invasive Detection of Microbial Growth in Media Fills
  • 本地全文:下载
  • 作者:David Brueckner ; David Roesti ; Ulrich Georg Zuber
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2016
  • 卷号:6
  • 期号:1
  • DOI:10.1038/srep27894
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:Two methods were investigated for non-invasive microbial growth-detection in intact glass vials as possible techniques for automated inspection of media-filled units. Tunable diode laser absorption spectroscopy (TDLAS) was used to determine microbially induced changes in O2 and CO2 concentrations within the vial headspaces. Isothermal microcalorimetry (IMC) allowed the detection of metabolic heat production. Bacillus subtilis and Streptococcus salivarius were chosen as test organisms. Parameters as robustness, sensitivity, comparability and time to detection (TtD) were evaluated to assess method adequacy. Both methods robustly detected growth of the tested microorganisms within less than 76 hours using an initial inoculum of <10CFU. TDLA turned out to be less sensitive than TDLA and IMC, as some false negative results were observed. Compared to the visual media-fill examination of spiked samples, the investigated techniques were slightly slower regarding TtD. Although IMC showed shorter TtD than TDLAS the latter is proposed for automating the media-fill inspection, as larger throughput can be achieved. For routine use either TDLA or a combination of TDLA and TDLA should be considered. IMC may be helpful for replacing the sterility assessment of commercial drug products before release.