摘要:Haemodialysis effectively removes small solutes and smaller-sized middle molecules from the blood; however, the clearance of larger middle molecules, which have been associated with negative effects, is poor. The novel medium cut-off (MCO) dialysis membrane has larger pore sizes and a more open structure than other high-flux membranes, providing improved removal of larger middle molecules while retaining albumin. However, larger pore sizes may potentially increase permeability to pyrogens, including endotoxins and other bacterial contaminants, that could be present in the dialysis fluid. In this study, we tested the capacity of low-flux, high-flux, MCO and high cut-off dialyser membranes with different pore sizes to prevent pyrogens crossing from dialysate to the blood side in a closed-loop test system, differentiating among lipopolysaccharides, peptidoglycans and bacterial DNA using a toll-like receptor assay. Even though the bacterial contamination levels in our test system exceeded the acceptable pyrogen dose for standard dialysis fluid, levels of lipopolysaccharides, peptidoglycans and bacterial DNA in the blood-side samples were too low to identify potential differences in pyrogen permeability among the membranes. Our results suggest that MCO membranes are suitable for haemodialysis using ISO standard dialysis fluid quality, and retain endotoxins at a similar level as other membranes.