Vibrational and numerical evaluation of human incus mechanical properties.
Gyliene, V. ; Kraptaviciute, N. ; Lipinski, P. 等
Vibrational and numerical evaluation of human incus mechanical properties.
1. Introduction
The mechanism of hearing involves conduction of mechanical
vibrations along the ossicular chain to the inner ear [1]. The middle
ear system with its smallest bones and articulations is responsible for
catching noise energy and transforming it to perilymph pressure changes
in the cochlea. The ossicular chain (the malleus, the incus and the
stapes - Fig. 1) connects the tympanic membrane and the inner ear, and
plays an important role in amplifying and regulating sound waves [2]. In
doing so the spatial motion of the ossicular chain appears as the one of
main causes of the hearing quality [3].
The degree of ossicular chain deterioration or damage in humans, as
a result of inflammatory disease, trauma or malformation, is variable
[5]. Of the three ossicles from the middle ear, the incus was often
reported in literature to be the most eroded bone and the more distant
from the inflammation focus [6, 7]. Borgstein et al [8] highlighted,
that the problem of erosion of the incus is a frequently observed, but
rarely discussed problem. Park [9] also examined the problems of the
long process of incus but caused by congenital origin. Finally, Yuang
[10] summarised ossicular damage in the patients with atelectasis and
found that for 72% of cases in posterior retraction pocket only incus
was eroded.
There are many ways to reconstruct the ossicular chain, but the
results are not always satisfactory because it is being done usually by
columella (bird) effect, which sometimes is unsatisfactory for hearing
impairment. Sometimes, some artificial materials, as glue [11] for
example, could be used to reconstruct the ossicular chain discontinuity.
Or the use of glass ionomer cement to repair incus long process defects
is a suitable method that improves hearing in pediatric patients [12].
It is hypothesised that the function of the malleus-incus-stapes
arrangement is to link the drum to the oval window with the flexibility
required for impedance matching but the rigidity to prevent
unconstrainable resonances from occurring in the hearing range [1]. If
it is true, then the structural stiffness of ossicular chain is the
critical design for middle-ear replacement prostheses.
Nevertheless, the substituting materials should fit according to
the material characteristics; the geometrical requirements are also
important. Kaftan et al [5] accentuate that the most difficult situation
is to evaluate the length of prosthesis. For this kind of problems, and
others, the finite element method (FEM) has distinct advantages in
modelling complex biological systems when compared to other techniques
[13]. Even, as reported in [14], it could improve the clinical surgical
intervention.
Consecutively to ossicles' surgery, the surgeons are observing
that incus bone work does not depend only on degradation level and
affected area size. It was also observed that sometimes damaged bone
part is large but it does not affect bone work in whole hearing system
and opposite--the bone degradation level can be low but the ossicles not
functioning well. From these observations one could make a hypothesis
that the bone work is mainly affected by the bone microstructure and
mechanical properties. Also, it can suppose that location of the focus
of infection affects bone strength and its ability to transmit
vibrations.
To confirm or deny this hypothesis a study of the middle ear bone
incus microstructure and mechanical properties was undertaken.
To identify bone microstructure [mu]CTomograph was used to evaluate
bone porosity and to identify bone quality. In order to find actual
value of incus bone Young's modulus, the modal analysis using
piezoelectric system was performed. Accordingly, it was decided to
extract the average bone properties by inverse method based on the
eigenvalue measurements and calculations. Consequently, FE model was
composed for selected incus sample.
2. Incus testing procedures
In total, four incus bones were prepared for this study. Each incus
bone was removed due to surgical treatment. Also, each sample was coded
for the privacy of patients. After using microscope with magnification
of 20 all samples were visually inspected to identify their condition.
It was found that the less damaged was the sample coded as
'1M' (sex of patient: women) (Fig. 2).
After visual inspection with optical microscope, all samples were
left to dry at room temperature and controlled humidity. The high
precision balance Sartorius CPA225D was used to perform the mass
evaluations. The mass of sample 1M was measured immediately after
removal from the refrigerator and from formaldehyde solution. The
results indicate that the mass variation for sample 1M was about 0.2%.
From the line-chart (Fig. 3) we can see that the sample mass has no
clear tendency to increase or to decrease during 98 hours period.
It could be said, that the mass changes are mainly due to the
ambient moisture variations. It can be concluded, that keeping samples
in the formaldehyde liquid does not affect their mass. It could mean
that incus bones are compact tissues.
Before tomography test each sample was glued to the plastic rod,
which served as a support during scanning. The X-ray measurements were
performed using "nanotom180N" device produced by GE Sensing
& Inspection Technologies phoenix|X-ray Gmbh. During the measurement
the tungsten target was used. The working parameters of X-ray tube were
V = 60 kV and I = 310 [micro]A. 1500 projections were taken with an
exposure time of 500 miliseconds with 4 integrations for each
exposition. The total time of measurement was around 60 min. The
reconstruction of measured objects was done with the aid of proprietary
GE software datosX ver. 2.1.0 with use of Feldkamp algorithm for cone
beam X-ray CT [15]. The final resolution of reconstructed object was 3.5
[micro]m.
The tomography analysis results showed that porosity in incus bone
is variable in its different sites testifying the highly heterogeneous
character of the bone mechanical properties (Fig. 4).
After scanning, the volume of each sample was next calculated using
image processing software Fiji[R]. Firstly, from all scanned slides we
eliminated slides including aluminium particle and support. Binariesed
data was then prepared consisting only of white and black pixels.
Finally, threshold values were set enabling the calculation of volumes
of sample. The volume of sample 1M was calculated twice: firstly, as for
the others samples, using scan results and later for the same model but
with manually eliminated porosity. This was done in order to compare
volumes and validate the FE model.
Table 1 presents the results from [mu]CTomography. It can be
summarized, that thanks to the tomography analysis results, the complex
microstructure of the sample was found and its volume and density was
precisely calculated with and without the sample porosity.
Various experimental methods can be used to identify the mechanical
properties of materials or tissues. The classical mechanical tests such
as traction, compression or torsion are frequently used to characterize
the properties of materials such as metals or polymers. However, they
require the sufficient amount of matter to machine the standard samples.
Moreover, they belong to the so-called destructive tests. Recently,
nano-indentation has been recognized as a powerful method to estimate
local values of Young's modulus of the materials [16, 17].
As the classical tests mentioned above, this method leads to a
local deterioration of the tested material. Also, it only provides the
local information concerning the elastic properties of the material
tested which can strongly change from site to site of the substantially
porous and inhomogeneous incus bone. On the other hand, ultrasonic
methods based on the measurement of sound wave celerity provide
convenient way to characterize the elastic properties of materials and
tissues in non-destructive manner. However, the small size of the
ossicle proscribes the use of the conventional ultrasonic devices.
Consequently, it was decided to extract the average bone properties by
inverse method based on the eigenvalues measurement and calculation.
Accordingly, the FE model of the incus 1M was built for Young Modulus
identification.
3. Modal analysis
To find real value of the average incus bone Young's Modulus
we decided to perform a modal analysis using piezoelectric exciter. The
experimental test bed was equipped with OPTIKA[R] TC B5 camera, with
resolution of 5Mpixels, and TOE7404 function generator, with frequency
range from 50Hz to 5MHz. All equipment used for this experimentation is
presented in Fig. 5.
Two samples (1VS and 1M) were simultaneously glued in the central
part of the piezoelectric plate (Fig. 6). One of them was the case study
sample--1M. Malleus-joint surface was chosen as a fixation point. After
gluing, samples were left to dry the glue for a few minutes. Next, using
function generator, the vibration frequency value was increased slowly
from 0 to 30 000 Hz.
To see better the body motion and to easily identify frequency
range for each mode, camera was zoomed only to the end of the most
flexible incus part, i.e. its long process (Fig. 7).
The registered frequency values are summarised in Table 2. As it
could be seen from Table 2, only three eigenvalues were obtained for
each sample. Two reasons can be mentioned to justify this low number.
Firstly, the used piezoelectric plate was not powerful enough to sweep
higher frequencies. Secondly, the inspected and found eigenfrequency
values concern only the incus long process.
4. Identification of Young Modulus by FE modelling
After the [mu]CTomography scanning the 3D geometry of incus was
"repaired" and meshed. Altair HyperMesh[R] Software was used
to close all surface holes and to make the 3D numerical model smooth. To
verify the convergence of the results two types of tetrahedral elements
(linear and quadratic) were used. Finally, the quadratic (10 nodes)
element type was selected for numerical simulation (Fig. 8).
After the meshing, 3D model was exported to MSC Marc[R] Software.
The initial material mechanical properties were chosen with reference to
literature [18]. We have chosen the value of 1.41 x [10.sup.10] Pa for
its Young's Modulus and 0.28 as the Poisson's ratio. As it was
found in the literature [13, 19] but also as a results of our tomography
analysis, the porosity or internal architecture of bone is not
homogeneous and differs from part to part of the bone.
Consequently, the density of the incus is not the same in the
different parts also. It means that the bone has different mechanical
properties in every part. For this reason, as it can be seen in Fig. 9
the FE model was divided in three parts--body (in pink), short process
(SP--in yellow) and long process (LP--in red).
Concerning the boundary conditions, three neighbouring nodes on
malleus-joint surface were constrained in X, Y, Z directions as
materialized by arrows in Fig. 9. As a first verification, the volume of
FE model was calculated using the MSC Marc[R] Software. The result
showed that volume of 1M-sample model was equal to 14.220 [mm.sup.3]. If
we compare this value with the results from [mu]CTomography analysis for
the same sample 1M taking into account the specimen porosity (V = 13.240
[mm.sup.3]), it clearly appears that the finite element mesh volume is
bigger than that calculated with Fiji[R] software.
However, if we compare the same mesh associated volume with 1M
sample considered as a continuum body (V = 14.163 [mm.sup.3]) it appears
that the difference is only 0.132 [mm.sup.3], i.e. less than 0.9%.
Because the volume difference between FE model and 1M-sample without
pores is very small, it could be concluded, that our 3D FE model is
sufficiently precise for numerical analysis.
To identify elastic properties of our sample we decided to deduce
the Young's modulus value by fitting results from vibration test.
3D FE model of 1M-sample with quadratic element type 184 of Marc library
(Full Integration & Assumed Strain) was used in order to extract the
eigenfrequencies of the sample. 10 calculations were done by changing
Young Modulus to fit the experimentally determined eigenfrequencies. The
calculations started with the value of Young Modulus equal to 1.41 x
[10.sup.10] Pa [18].
The resulting eigenfrequencies showed that the Young's Modulus
equal to 4.5 x [10.sup.+09] Pa provided the response closest to the
experimentally obtained eigenvalues. Also, it should be mentioned that
these calculations were made using bone model in which porosity was not
included. Secondly, the inhomogeneity in the FE model was introduced.
According to the literature [13, 19, 20] the eigenvalue calculations
were performed with different value of long process density equal to
5080 kg/[m.sup.3] comparing to body density of 2030 kg/[m.sup.3]. By
doing this, it was found that when the density of our sample is not
homogeneous the frequency values obtained for E = 4.5 x [10.sup.+09] Pa
are lower than experimental ones. Consequently, the Young's Modulus
was increased to E = 6 x [10.sup.+09] Pa, enabling us to fit better the
experimental values.
5. Conclusions and discussions
In order to identify incus bone micro structure and mechanical
properties, firstly the mass of the sample was measured. It was noticed
that mass changes appeared because of the ambient humidity variation,
but not because of bone water contents. Bone volume and microstructure
analysis was performed using [mu]CTomograph From tomography results we
founded that microstructure and porosity of bone are not homogeneous and
differ from part to part of the bone. It could be said, that our samples
are more porous than typical cortical bone.
To find actual value of incus bone Young's Modulus, the
experimental modal analysis was made. Due to piezoelectric plate power
problems and results registration method based on the visual inspection
of sample motion in one direction, only three frequency values were
identified. Comparing frequency values obtained from experiment with
frequencies calculated by changing Young's Modulus value, it was
found that the calculated eigenfrequencies are closest to real values if
the Young's Modulus of the incus is equal to 4.5 x [10.sup.+09] Pa.
However the obtained results are only indicative, because calculations
were made not taking into account the bone porosity.
Moreover, on purpose to find out incus bone mechanical properties,
the real density of each bone part has to be identified. Also, to get
reliable results, all calculations should be repeated using bone model
with porosity. Furthermore, it is important to mention that, to get
reliable results for this research more than one computational model
should be used.
Acknowledgements
The preparation of samples was performed fulfilling the
requirements of ethical committee of Lithuanian University of Health
Sciences.
References
[1.] Ferris, P.; Prendergast, P.J. 2000. Middle-ear dynamics before
and after ossicular replacement, J. Biomech. 33(5): 581-590.
http://dx.doi.org/10.1016/S0021-9290 (99)00213-4.
[2.] Stoessel, A.; Gunz, P.; David, R; Spoor, F. 2016. Comparative
anatomy of the middle ear ossicles of extant hominids - Introducing a
geometric morphometric protocol, J. Hum. Evol. 95: 1-25.
http://dx.doi.org/10.1016/jjhevol.2015.10.013.
[3.] Kovincic, N.I.; Spasic, D.T. 2016. Dynamics of a middle ear
with fractional type of dissipation, Nonlinear Dyn. 85(4): 2369-2388.
http://dx.doi.org/10.1007/s11071-016-2832-z.
[4.] Stropus, R.; Tamasauskas, K.A.; Pauziene, N. 2005. The human
anatomy [online, accessed 6 June 2016], Available from Internet:
http://ebooks.vitaelitera.lt/eb/1449/zmogaus-anatomija/(in Lithuanian).
[5.] Kaftan, H.; Bohme, A.; Martin, H. 2015. Parameters for novel
incus replacement prostheses, Eur. Arch. Otorhinolaryngol. 273(9):
2411-2417. http://dx.doi.org/10.1016/10.1007/s00405-015-3810-7.
[6.] Albera, R.; Canale, A.; Piumetto, E.; Lacilla, M.; Dagna, F.
2012. Ossicular chain lesions in cholesteatoma, Acta Otorhinolaryngol.
Ital. 32(5): 309-313.
[7.] Duboeuf, F.; Burt-Pichat, B.; Farlay, D.; Suy, P.; Truy, E.;
Boivin, G. 2015. Bone quality and biomechanical function: A lesson from
human ossicles, Bone. 73: 105-110.
http://dx.doi.org/10.1016/j.bone.2014.12.009.
[8.] Borgstein, J.; Gerritsma, T.V.; Bruce, I.A. 2008. Erosion of
the incus in pediatric posterior tympanic membrane retraction pockets
without cholesteatoma, Int. J. Pediatr. Otorhinolaryngol. 72(9):
1419-1423. http://dx.doi.org/10.1016/j.ijport.2008.06.004.
[9.] Park, K.; Choung, Y.-H.; Shin, Y. R.; Hong, S.P. 2007.
Conductive deafness with normal eardrum: absence of the long process of
the incus, Acta Otolaryngol. 127(8): 816-820.
http://dx.doi.org/10.1080/00016480601075449.
[10.] Yung, M.W. 1997. Retraction of the pars tensa long-term
results of surgical treatment, Clin. Otolaryngol. Allied Sci. 22(4):
323-326. http://dx.doi.org/10.1046/j.1365-2273.1997.00018.x.
[11.] Gungor, V.: Atay, G.; Bajin, M. D.; Yarali, M. Sarac, S.;
Sennaroglu, L. 2016. Comparison of various bone cement ossiculoplasty
techniques and functional results, Acta Oto-Laryngol. 136 (9): 883-887.
http://dx.doi.org/10.3109/00016489.2016.1172255.
[12.] Atan, D.; Dere, H.; Yamur, A.R; Ozcan, K.M. 2016 Results of
ossicular chain reconstruction with glass ionomer cement in pediatric
patients, Int. J. Pediatr. Otorhinolaryngol. 85: 103-106.
http://dx.doi.org/10.1016/j.ijporl.2016.03.027.
[13.] Sun, Q.; Gan, R.Z.; Chang, K.-H.; Dormer, K.J. 2002.
Computer-integrated finite element modeling of human middle ear,
Biomech. Model. Mechanobiol. 1(2): 109-122.
http://dx.doi.org/10.1007/s10237-002-0014-z.
[14.] Volandri, G.; Di Puccio, F.; Forte, P.; Carmignani, C. 2011.
Biomechanics of the tympanic membrane, J. Biomech. 44(7): 1219-1236.
http://dx.doi.org/10.1016/jjbiomech.2010.12.023.
[15.] Feldkamp, L.A.; Davis, L.C.; Kress, J.W. 1984. Practical
cone-beam algorithm, JOSA A, 1(6): 612-619.
[16.] Panich, N.; Yong, S. 2005. Improved method to determine the
hardness and elastic modulus using nanoindentation, KMITL Sci. Tech. J.
5(2): 483-492.
[17.] Oliver, W.C.; Pharr, G.M. 1992. An improved technique for
determining hardness and elastic modulus using load and displacement
sensing indentation experiments, J. Mater. Res. 7: 1564-1583.
[18.] Zhang, X.; Guan, X.; Nakmali, D.; Palan, V.; Pineda, M.; Gan,
R.Z. 2014. Experimental and modeling study of human tympanic membrane
motion in the presence of middle ear liquid, Jaro-J. Assoc. Res.
Otolaryngol. 15(6): 867-881.
http://dx.doi.org/10.1007/s10439-012-0624-2.
[19.] Liu, H.; Rao, Z.; Huang, X.; Cheng, G.; Tian, J.; Ta, N.
2014. An incus-body driving type piezoelectric middle ear implant design
and evaluation in 3D computational model and temporal bone, Sci. World
J. 121624:1-8. http://dx.doi.org/10.1155/2014/121624.
[20.] Lee, D.; Ahn, T.-S. 2015. Statistical calibration of a finite
element model for human middle ear, J. Mech. Sci. Technol. 29 (7):
2803-2815. http://dx.doi.org/ 10.1007/s12206-015-0609-9.
V. Gyliene, N. Kraptaviciute, P. Lipinski, S. Wronski, R. Rahouadj,
J. Tarasiuk, A. Baldit, G. Gylys, K. Norkaityte
VIBRATIONAL AND NUMERICAL EVALUATION OF HUMAN INCUS MECHANICAL
PROPERTIES
Summary
The capabilities to apply the FEM in hearing research field are
large: the identification of tissues properties, the definition of the
behaviour of hearing elements or even it could improve the clinical
surgical intervention. Numerous researchers are focused on hearing
improvement or understanding problems by working on middle ear or inner
ear behaviour.
In our work we focused on mechanical properties identification of
incus by using micro-CT scanning, performing vibrational analysis and
finally the FE simulations. Contrarily to other researchers we supposed
incus bone to be non-homogeneous body and accordingly attributed
different densities in long and short processes. The non-homogeneity of
the long and short processes could be the reason why the incus bone
behaviour does not depend simply on degradation level and affected area
size. Finally, by inverse FE modelling, the Young Modulus was estimated.
Keywords: Finite element modelling, mechanical properties, hearing,
incus.
Received October 27, 2016
Accepted December 2, 2016
V. Gyliene (*), N. Kraptaviciute (*), P. Lipinski (**), S. Wronski
(***), R. Rahouadj (****), J. Tarasiuk (***), A. Baldit (**), G. Gylys
(*****), K. Norkaityte (*****)
(*) Kaunas University of Technology, Studentu 56, LT-51424 Kaunas,
Lithuania, E-mail:
[email protected],
[email protected]
(**) University of Lorraine, ENIM - LaBPS, 1 route d'Ars
Laquenexy, FR-57078 Metz Cedex 3, France, E-mail:
[email protected],
[email protected]
(***) AGH University of Science and Technology, Al Mickiewicza 30,
PL-30059 Krakow, Poland, E-mail:
[email protected],
[email protected]
(****) University of Lorraine, LEMTA, 2 Ave Foret Haye, TSA 60604,
F-54518 Vandoeuvre Les Nancy, France, E-mail:
[email protected]
(*****) Lithuanian University of Health Sciences, Department of
Oto-Rhino-Laryngology, Eiveniu 2, LT-50009 Kaunas, Lithuania, E-mail:
[email protected],
[email protected]
Table 1
Incus properties, obtained from [mu]CTomography
Sample Bone volume, Mass, Density,
1M [mm.sup.3] g g/[mm.sup.3]
without pores 14.163 0.02881 0.002034
with pores 13.240 0.02881 0.002176
Table 2
Registered eigenfrequency values
Mode Number 1VS Sample (*) 1M Sample
1 18.93 kHz 14.18 kHz
2 19.96 kHz 20.67 kHz
3 38.81 kHz 30.00 kHz
(*) Remark: sample was not submerged in formaldehyde
COPYRIGHT 2016 Kauno Technologijos Universitetas