期刊名称:IOP Conference Series: Earth and Environmental Science
印刷版ISSN:1755-1307
电子版ISSN:1755-1315
出版年度:2018
卷号:121
期号:2
页码:022033
DOI:10.1088/1755-1315/121/2/022033
语种:English
出版社:IOP Publishing
摘要:Determining fluoride ions in ammonia desulphurization slurry using a fluoride ion selective electrode (ISE) is investigated. The influence of pH was studied and the appropriate total ionic strength adjustment buffer and its dosage were optimized. The impact of Fe3+ concentration on the detection results was analyzed under preferable conditions, and the error analysis of the ISE method's accuracy and precision for measuring fluoride ion concentration in the range of 0.5-2000 mg/L was conducted. The quantitative recovery of F− in ammonium sulfate slurry was assessed. The results showed that when pH ranged from 5.5~6 and the Fe3+ concentration was less than 750 mg/L, the accuracy and precision test results with quantitative recovery rates of 92.0%-104.2% were obtained.