  • 标题:A Case of SmartPlug Partial Extrusion through Canalicular Mucosa
  • 本地全文:下载
  • 作者:Choi, Hwa Su ; Lee, Seung Ki ; Park, Jong Seok
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2017
  • 卷号:58
  • 期号:4
  • 页码:459-462
  • DOI:10.3341/jkos.2017.58.4.459
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To report a case of a SmartPlug that partially extruded through the canalicular mucosa.

    Case summary

    A 62-year-old female patient visited our Department of Ophthalmology for redness and discharge in her right eye that began worsening a week prior. The patient was diagnosed with severe dry eye syndrome via Sjogren's syndrome. The SmartPlug was inserted into her left lower punctum 12 years prior, and another plug was inserted in the right 9 years prior. Nothing notable was found in her history. The best corrected visual acuity and intraocular pressure at the initial visit was 1.0 and 14 mmHg, respectively, in both eyes. Swelling, erythema, and tenderness at the right lower punctum were identified. As we irrigated the lower lacrimal system, partial obstruction was suspected and a large amount of mucopurulent discharge in the conjunctival fornix with conjunctival injection was seen. Under the right lower punctum, a 1 × 1 mm yellowish foreign body was partially extruding from the canalicular mucosa. The foreign body was removed, and a round defect was left. The patient's symptom completely resolved after 4 weeks of topical antibiotics and ointment treatment.

    Conclusions

    In patients who have a SmartPlug inserted prior to visits for redness and discharge, a SmartPlug plug extrusion should be considered.

  • 关键词:Complication; Extrusion; Punctal plug; SmartPLUG
国家哲学社会科学文献中心版权所有