首页    期刊浏览 2024年12月02日 星期一
登录注册

文章基本信息

  • 标题:Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry
  • 本地全文:下载
  • 作者:Park, Ae-Ja ; Yoo, Jun-Il ; Choi, Jee-Hye
  • 期刊名称:Journal of Bone Metabolism
  • 印刷版ISSN:2287-6375
  • 出版年度:2017
  • 卷号:24
  • 期号:2
  • 页码:105-109
  • DOI:10.11005/jbm.2017.24.2.105
  • 语种:English
  • 出版社:The Korean Society for Bone and Mineral Research
  • 摘要:Background

    The purpose of this study was to calculate the measurement uncertainty of the process of bone mineral density (BMD) analysis using dual energy X-ray absorptiometry with traceability.

    Methods

    Between March 2015 and October 2016, among healthy participants in their 20s and 30s, the study included those who had not taken calcium, vitamin D supplements and steroids and were without a history of osteoporosis, osteopenia and diseases related to osteoporosis. Relational expression of the model was established based on Guide to the Expression of Uncertainty in Measurements and Eurachem and the uncertainty from each factor was evaluated.

    Results

    The combined standard uncertainty was 0.015, while the expanded uncertainty was 0.0298. The factor-specific standard uncertainties that occurred in the process of measuring BMD were 0.72% for the calibration curve, 0.9% for the internal quality control (IQC) using Aluminum Spine Phantom, 0.58% for European Spine Phantom (ESP), and 0.9% for the inspector precision (IP).

    Conclusions

    The combined standard uncertainty of the spine BMD corrected with ESP was 0.015 when measured at one time and targeting one participant. The uncertainties of the accuracy of the IQC and the IP were higher than that of the other factors. Therefore, there will be a need for establishment of protocols to lower these uncertainties.

  • 关键词:Absorptiometry photon; Bone Density; Lumbar Vertebrae; Uncertainty