出版社:American Society for Biochemistry and Molecular Biology
摘要:We compared two HPLC methods (anion exchange [AE] and steric exclusion [SE]) for analysis of mouse lipoprotein profiles by determining coefficients of variability (CVs) under varying conditions. CVs for AE and SE were comparable on fresh samples. There was an inverse relationship between subfraction curve area and CV [ r = −0.65 (AE) and −0.50 (SE)], consistent with the interpretation that as curve area decreased, error variance increased and signal-to-noise ratio decreased. Sample storage did not affect SE. In contrast, with AE, alterations in measured lipoproteins were apparent after storage, including a decrease in the HDL subfraction [66.8% (baseline) vs. 15.9% (1 week); P r = 0.69 (AE) and 0.87 (SE)]. Both AE and SE yield reproducible, accurate, and rapid measurements of lipoproteins from small amounts of serum. AE yields more sensitive, high-amplitude, well-defined peaks that can be easily distinguished and necessitates the use of smaller sample volumes compared with SE, but sample storage causes alterations in the chromatogram. SE appears better suited to serial analyses involving stored samples.