期刊名称:International Journal of Multimedia and Ubiquitous Engineering
印刷版ISSN:1975-0080
出版年度:2016
卷号:11
期号:6
页码:27-34
DOI:10.14257/ijmue.2016.11.6.03
出版社:SERSC
摘要:In order to reconstruct the 3D temperature field based on the binocular vision system, we carry out a research on the calibration of the infrared camera. To start with, we make a calibration board which can be clearly identified by the infrared camera; then, with the help of OpenCV, we implement the calibration of two infrared cameras respectively and rectify the infrared images. According to the parameters obtained from the single infrared camera calibration, we accomplish the stereo calibration. The experimental results show that the average reprojecting error is below 0.14 pixels and when using the system to carry out a real measurement, the average error is 3.12%. The feasibility of the proposed calibration approach is verified, the experimental measurement results is close to the ground truth.
关键词:3D temperature field; Vision measurement; Infrared camera calibration; ; Infrared image matchup