首页    期刊浏览 2024年11月30日 星期六
登录注册

文章基本信息

  • 标题:Novel Self-inflating Resuscitators Maintaining High FiO2 with High Minute Ventilation
  • 本地全文:下载
  • 作者:Lee, Jeong Rim ; Yoo, Seung Yeon ; Shin, Teo Jeon
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2005
  • 卷号:48
  • 期号:3
  • 页码:232-234
  • DOI:10.4097/kjae.2005.48.3.232
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:Background

    Self inflating resuscitators are widely used for the patients who need positive pressure ventilation during transport. During self inflating resuscitator ventilation, the FiO2 may decreases even with the use of the oxygen reservoir. Such phenomenon may increase the difficulty and risk of transport of the critically ill patients. Twin self inflating resuscitator was assembled by being modified from two conventional ones to achieve more stable FiO2.

    Methods

    The twin self inflating resuscitator and one conventional adult self inflating resuscitator were tested on a test lung. In the twin resuscitator, two adult self inflating resuscitators were connected serially without distal unidirectional valve plate. The resuscitators were compressed at variable tidal volumes, respiratory rates and oxygen flow rates, guided by the monitor.

    Results

    With conventional self inflating resuscitator ventilation, the FiO2 was maintained over 95% until minute ventilation of 7,500, 1,4000, 17,500 ml at respectively 5, 10, 15 L/min oxygen flow rate. With serial type, the FiO2 started to decrease with the minute ventilation over 12,500, 24,000, 28,000 ml at 5, 10, 15 L/min oxygen flow rate.

    Conclusions

    By simple connection of two self inflating resuscitators, the FiO2 during self inflating resuscitator ventilation could be maintained during almost two times of minute ventilation compared to conventional ones.

  • 关键词:intrahospital transport; self-inflating resuscitator
国家哲学社会科学文献中心版权所有