  • 标题:Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study
  • 本地全文:下载
  • 作者:Frédéric Duprez ; Maria Barile ; Thierry Bonus
  • 期刊名称:Health
  • 印刷版ISSN:1949-4998
  • 电子版ISSN:1949-5005
  • 出版年度:2014
  • 卷号:06
  • 期号:15
  • 页码:1978-1983
  • DOI:10.4236/health.2014.615232
  • 语种:English
  • 出版社:Scientific Research Publishing
  • 摘要:The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient’s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low flow rate was required. Compensated-pressure oxygen flowmeters for these low rates were more accurate than their non-compensated counterparts. This study emphasizes the need to individually adapt the oxygen flow rate each time a patient has to move from one flowmeter to another.
  • 关键词:Medical Device Accuracy; Oxygen Therapy; Flowmeter