摘要:Abstract
Background
Hypertrophic scarring after burn injury is one of the greatest unmet challenges in patients with burn injuries. A better understanding of the characteristics of scar maturation and early prediction of the long-term outcome of scarring are prerequisites for improving targeted therapies and pivotal for patient counselling.
Methods
Repeated measurements of scar stiffness in 11 pediatric patients were performed over the course of 1 year using 2 suction devices, the Cutometer and the Nimble. In addition, the observer pliability score of the Patient and Observer Scar Assessment Scale was applied. This longitudinal study allowed quantification of the ability of each of the measured parameters to reflect scar maturation, as indicated by change in skin pliability/stiffness, over time (using linear regression); the ability to distinguish individual patients (intraclass correlation coefficient (ICC)); the correlation of the devices (Spearman correlation coefficient); and the ability to predict long-term scar maturation based on early scar assessment (using receiver operating characteristic).
Results
All the tools used showed significant longitudinal decrease of scar stiffness from 3 months until 12 months after the injury. The Nimble (ICC
patient
Nimble = 0.99) and the Cutometer (ICC
patient
Cuto = 0.97) demonstrated an excellent ability to distinguish between individual patients. The Nimble seemed to be able to predict the 12-month pliability of scars based on early (3-month) measurements (area under the curve (AUC)
12m
POSAS = 0.67; AUC
12m
C = 0.46; AUC
12m
N = 0.79).
Conclusions
The results of this preliminary study suggest that all 3 tools provide suitable means to quantify alterations in scar stiffness over time. Initial evidence suggests the Nimble is most favorable for predicting changes in stiffness associated with long-term scar maturation. Further studies with a larger sample size are required to validate tissue suction as a clinical tool for analysis of changes of scar stiffness over time.