摘要:AbstractBackgroundWith the application of the scanning transmission electron microscopy (STEM) detector, transmitted electron images similar to transmission electron microscopy (TEM) can be obtained from scanning electron microscopy (SEM), which is referred to STEM-in-SEM imaging. Compared to TEM-energy dispersive spectroscopy (EDS), SEM-EDS is expected to be more efficient and reliable for chemical analysis of TEM specimens due to the larger sample space in SEM and the higher take-off angle of the SEM-EDS detector. Unfortunately, this advantage is not well applied to TEM specimens in practice, mainly because of fault signals generated from the commercial grid holders used in SEM and STEM-in-SEM. This study is designed to solve this problem by modifying the commercial holders and to test them through EDS analysis of apatite phases.FindingsWe first changed the way of assembling parts of the commercial multi-grid holder for SEM. This new assembly was capable of producing a thinner upper cover part, resulting in the reduction of the EDS fault signals generated from the holder. Furthermore, the thin upper part allowed us to get images in shorter working distances, that is, at higher magnifications. This design concept was also applied to the commercial single-grid holder for STEM-in-SEM, producing a new multi-grid holder to be used for loading multiple samples.ConclusionWe confirmed that the new holders produced reliable chemical data from apatite phases. In case of oxygen analysis, despite of low electron brightness from the tungsten source, 5 kV provided more stable acquisition and signal yields than 15 kV. We expect that these modified holders facilitate more efficient EDS analysis for multiple samples under the same analytical conditions in SEM and STEM-in-SEM.