摘要:AbstractThe article deals with mathematical modelling of a system for ex-vivo lung perfusion. The system which is modelled here is named Mobile Circulatory Module and is designed for transport of live lungs namely porcine or human lungs. Transported lungs should be used for various research or even for transplantation. The article describes mathematical model of thermal circuit and hydraulic circuit which are vital parts of the MCM to secure living conditions of the transported organ. Mathematical model is simulated with MATLAB Simulink and shows good qualitative correspondence with real behavior of the system. Based on math model there is designed control algorithm for temperature control and there is discussed control of perfusion liquid as well.
关键词:Keywordsex-vivo lungs perfusionisolated organMCMMatlab modelcontrol of flowtemperature control